- 13 -
sample Same as menu File->Save->Sample (to disk)
Save dark [picture of DARK light bulb]
spectrum Used for Absorbance, Transmittance, and
Irradiance modes.
Left Click here with light off.
Caution: Right Click will zero previous dark.
Save reference [picture of YELLOW light bulb]
spectrum Used for Absorbance and Transmittance modes.
Left Click here with light on and
without sample in place.
When viewing Irradiance -> watts per square
meter, clicking here starts the UV monitor
application.
Print graph [picture of printer with white paper]
Left Click to plot graph on printer.
Snapshot spectra [picture of camera]
Left Click here to freeze graph trace.
Left Click here again to unfreeze!
Zoom wavelength [picture of Z]
Reminds you how to X and Y Zoom.
Left Click here again to UN-Zoom! Also,
Left Click on left side of graph to UN-Zoom.
Right Click to enable Y zoom mode.
Re-scale Y axis [picture of graph with up+down arrows]
Left Click here to AUTOSCALE Y axis.
Right Click to UNDO AUTOSCALE Y axis.
Also use View->Y set scale menu to
override AUTOSCALE with selected scale.
Compute Area [picture of peak area with green]
Left Click here to show the AreaPSD,
Centroid, Peakwave, FWHM, and Centbase
of a peak. If the data cursor is
not on a peak, the icon will seek to a
peak in the closest direction.
Right Click to start dual data cursor
measurement as outlined below.
This information is printed in the graph
title (concatenated to end of line) when
printing a graph.
Compute Area toolbar function details
-------------------------------------
1. Use left mouse click to compute and display:
AreaPSD, Centroid, Peakwave, FWHM, and Centbase
>> Automatically selects closest peak & finds baseline.
>> AreaPSD = integral Area (Power Spectral Density).
>> Centroid = center wavelength of AreaPSD in nm.
>> Peakwave = wavelength of the tallest point.
>> FWHM = Full Width Half Max of peak in nm.
Kommentare zu diesen Handbüchern